Design Control is one of the most important quality
elements of medical device quality system. Table below tabulate each element of
design control(1,2). I just copied verbatim directly from the FDA website.
Since the brevity of the blog post, it is impractical to
discuss every element in detail. But,
when initiating a medical device, follow and comply with all elements of design control. The keys are having a plan, write procedures for everything and documents every activities in detail about the medical device. Have all the protocols and reports of all testing filed and organized in one location. Always keep the Design History File (DHF) most updated.
when initiating a medical device, follow and comply with all elements of design control. The keys are having a plan, write procedures for everything and documents every activities in detail about the medical device. Have all the protocols and reports of all testing filed and organized in one location. Always keep the Design History File (DHF) most updated.
Basically, Design Controls are an interrelated set of practices
and procedures that are incorporated into the design and development process
(2).
Subpart
C: 21 CFR 820.30 Design Control(1)
|
|
820.30 Design
Controls
(a) General |
(1) Each manufacturer of any class III or class II device,
and the class I devices listed in paragraph (a)(2) of this section, shall
establish and maintain procedures to control the design of the device in
order to ensure that specified design requirements are met
|
21 CFR 820.30
Design Control (b) Design and development planning.
|
Each manufacturer shall establish and maintain plans that
describe or reference the design and development activities and define
responsibility for implementation. The plans shall identify and describe the
interfaces with different groups or activities that provide, or result in,
input to the design and development process. The plans shall be reviewed,
updated, and approved as design and development evolves.
|
21 CFR 820.30
Design Control (c)Design input.
|
Each manufacturer shall establish and maintain procedures to
ensure that the design requirements relating to a device are appropriate and
address the intended use of the device, including the needs of the user and
patient. The procedures shall include a mechanism for addressing incomplete,
ambiguous, or conflicting requirements. The design input requirements shall
be documented and shall be reviewed and approved by a designated
individual(s). The approval, including the date and signature of the
individual(s) approving the requirements, shall be documented.
|
21 CFR 820.30
Design Control (d)Design output
|
Each manufacturer shall establish and maintain procedures
for defining and documenting design output in terms that allow an adequate
evaluation of conformance to design input requirements. Design output
procedures shall contain or make reference to acceptance criteria and shall
ensure that those design outputs that are essential for the proper
functioning of the device are identified. Design output shall be documented,
reviewed, and approved before release. The approval, including the date and
signature of the individual(s) approving the output, shall be documented.
|
21 CFR 820.30
Design Control (e)Design review.
|
Each manufacturer shall establish and maintain procedures
to ensure that formal documented reviews of the design results are planned
and conducted at appropriate stages of the device's design development. The
procedures shall ensure that participants at each design review include
representatives of all functions concerned with the design stage being
reviewed and an individual(s) who does not have direct responsibility for the
design stage being reviewed, as well as any specialists needed. The results
of a design review, including identification of the design, the date, and the
individual(s) performing the review, shall be documented in the design
history file (the DHF).
|
21 CFR 820.30
Design Control (f)Design
verification.
|
Each manufacturer shall establish and maintain procedures
for verifying the device design. Design verification shall confirm that the
design output meets the design input requirements. The results of the design
verification, including identification of the design, method(s), the date,
and the individual(s) performing the verification, shall be documented in the
DHF.
|
21 CFR 820.30
Design Control (g)Design
validation.
|
Each manufacturer shall establish and maintain procedures
for validating the device design. Design validation shall be performed under
defined operating conditions on initial production units, lots, or batches,
or their equivalents. Design validation shall ensure that devices conform to
defined user needs and intended uses and shall include testing of production
units under actual or simulated use conditions. Design validation shall
include software validation and risk analysis, where appropriate. The results
of the design validation, including identification of the design, method(s),
the date, and the individual(s) performing the validation, shall be
documented in the DHF.
|
21 CFR 820.30
Design Control (h)Design
transfer.
|
Each manufacturer shall establish and maintain procedures
to ensure that the device design is correctly translated into production
specifications.
|
21 CFR 820.30
Design Control (i) Design Changes
|
Each manufacturer shall establish and maintain procedures for
the identification, documentation, validation or where appropriate
verification, review, and approval of design changes before their
implementation.
|
21 CFR 820.30
Design Control (j)Design
history file.
|
Each manufacturer shall establish and maintain a DHF for
each type of device. The DHF shall contain or reference the records necessary
to demonstrate that the design was developed in accordance with the approved
design plan and the requirements of this part.
|
2.
–Design
Control Guidance For Medical Device Manufacturers, March 11 1997, FDA
Disclaimer: Although
the author had exhaustively researched all sources to ensure the accuracy and
completeness of the information contained in this blog, but no warranty and
fitness is implied. I assumed no responsibility and implied warranty of any
kind for errors, inaccuracies, omission, or any inconsistency herein. No
liability is assumed for incidental or consequential damages in connection with
the use of the information contained herein. Readers should always use their
own judgment and review all related regulatory guidelines. Guidelines can
change over time.
No comments:
Post a Comment